Model Number 8888160341
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code Hemorrhage/Blood Loss/Bleeding (1888)
Date of Event 08/19/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An investigation is currently underway.Upon completion, the results will be forwarded.
 
Event or Problem Description
The customer reported that the umbilical vessel catheter (uvc) had a hole in it and was leaking.They placed it in the patient on (b)(6) 2022 and saw that blood was dripping from the tubing, not a connector.There was no significant blood loss and the device was in place for less than a day.A new uvc was replaced immediately after initial insertion without any additional complications.However, the patient underwent additional x-ray exposure.
 
Additional Manufacturer Narrative
A review of the device history records (dhr) yielded no discrepancies within the manufacture of the affected lot number reported.All quality assurance testing performed during manufacturing was acceptable.The quality assurance review of the visual, physical, and dimensional evaluation results indicated that the product met specification requirements.In addition, all dhrs are reviewed to confirm all regulatory requirements prior to product release.One used catheter from lot number 1803800115 and product code id 8888160341 was received inside of a generic plastic bag with its original sealed package.According to the test performed on the device, the catheter did showed signs of use such as blood residue.The physical sample evaluation revealed that the sample passed the leak test during underwater testing, therefore, no deficiency was identified.The customer reported event was not confirmed based on the visual and functional evaluation.A change was made to change of design of the catheter and a formal corrective/preventative action was opened after the reported lot was manufactured.Customer perception is determined as the most probable root cause based on the investigation as there are controls in place to detect this type of condition such as leak test and 100% occlusion tests.Based on the available information, it can be concluded that product was manufactured according to specifications.It must be noted that in-process controls such as personnel training, incoming quality acceptance testing for raw material, 100% in process visual inspection and visual acceptance sampling are performed in the plant are in place to prevent nonconforming product from leaving the manufacturing operations.This complaint will be used for tracking and trending purposes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UMBILICAL CATHETER SINGLE LUMEN 5FR
Common Device Name
CATHETER, UMBILICAL ARTERY
Manufacturer (Section D)
CARDINAL HEALTH
777 west street
mansfield MA 02048
Manufacturer (Section G)
CARDINAL HEALTH
edificio b20 calle #2 zona fra
alajuela
CS  
Manufacturer Contact
jill saraiva
777 west street
mansfield, MA 02048
5086183640
MDR Report Key15422249
Report Number3009211636-2022-00793
Device Sequence Number11459511
Product Code FOS
UDI-Device Identifier20884527005120
UDI-Public20884527005120
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number8888160341
Device Catalogue Number8888160341
Device Lot Number1803800115
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/22/2022
Supplement Date Received by Manufacturer08/22/2022
Initial Report FDA Received Date09/14/2022
Supplement Report FDA Received Date10/17/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-