| Model Number |
9438-06 |
| Medical Device Problem Code |
Wireless Communication Problem (3283)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/23/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that signal loss over one hour occurred.Data was received for evaluation.However, the alleged product is not present within the investigation window.The allegation and a probable cause could not be determined.No injury or medical intervention was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that signal loss over one hour occurred.The product was evaluated.An external visual inspection was performed and passed.Perform voltage test was performed and passed.Nordic bluetooth pairing test was performed and passed.A review of the bin file was performed and signal loss was not found within the investigation window.However, the device operated within specification.The allegation was not confirmed.The probable cause could not be determined.No injury or medical intervention was reported.
|
| |
|