• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VERAN MEDICAL TECHNOLOGIES, INC ALWAYS-ON TIP TRACKED 21GA ANSO CYTOLOGY NEEDLE; BIOPSY NEEDLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VERAN MEDICAL TECHNOLOGIES, INC ALWAYS-ON TIP TRACKED 21GA ANSO CYTOLOGY NEEDLE; BIOPSY NEEDLE Back to Search Results
Model Number INS-0392
Medical Device Problem Code Retraction Problem (1536)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/18/2022
Type of Reportable Event Malfunction
Event or Problem Description
21ga needle was fully extended out of the packaging and would not retract.
 
Additional Manufacturer Narrative
At the time of this report, the device had not been returned for investigation.Device was not used during the procedure.
 
Additional Manufacturer Narrative
This report is being supplemented to provide information that was inadvertently not included in the initial medwatch report, and to provide the results of the investigation.G3: date received by manufacturer was blank in the initial medwatch report and should have been 18-aug-2022.The subject device was not returned; therefore, there was insufficient evidence to confirm the failure mode.The possible failure modes associated with an unretractable needle tip include damage to the handle, damage to the push rod, damage to the pullwire or the pullwire-needle junction.Manufacturing or assembly errors involving the length of key components, including the pullwire or the sheath, could affect the needle's ability to extend and retract.Veran will continue to monitor field performance for this device.This supplemental report is being submitted as a corrective action following a retrospective review of complaints in response to an observation from an external inspection.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALWAYS-ON TIP TRACKED 21GA ANSO CYTOLOGY NEEDLE
Common Device Name
BIOPSY NEEDLE
Manufacturer (Section D)
VERAN MEDICAL TECHNOLOGIES, INC
1938 innerbelt business center
saint louis MO 63114
Manufacturer (Section G)
VERAN MEDICAL TECHNOLOGIES, INC
1938 innerbelt business center
saint louis MO 63114
Manufacturer Contact
jennifer andre
1938 innerbelt business center
saint louis, MO 63114
3146598500
MDR Report Key15422493
Report Number3007222345-2022-00020
Device Sequence Number14092301
Product Code JAK
UDI-Device Identifier00815686020613
UDI-Public00815686020613
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K170023
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/09/2023
Device Model NumberINS-0392
Device Catalogue NumberINS-0392
Device Lot Number04619200831
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer02/09/2023
Initial Report FDA Received Date09/14/2022
Supplement Report FDA Received Date08/08/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/20/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-