Per regulation eu (b)(4) (general data protection regulation), patient identifiers were not collected or recorded and therefore are not available.If implanted, give date: not applicable, as the lens was not implanted.If explanted, give date: not applicable, as the lens was not implanted; therefore, it was not explanted.Telephone number: (b)(6).The intraocular lens (iol) has not returned for evaluation; therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
|
|
Additional information: section d9: device available for evaluation? yes.Returned to manufacturer on: 22-sep-2022.Section h3: device evaluated by manufacturer? yes.Device evaluation: visual inspection under magnification revealed that the lens was received with both haptics detached and bent.The lens was cleaned and two tears on the iol were identified.A scratch on the optic body was also observed.The complaint issue was identified during product evaluation; however, based on the complaint investigation results the complaint issue could not be confirmed to be related to the manufacturing or design process.The other observed issues during the product evaluation could not be confirmed to be related to a manufacturing or design issue.Therefore, there is no indication of a product deficiency or product malfunction.Conclusion: as a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
|