ST PAUL BCI CAPNOCHECK PLUS; ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE Back to Search Results
Model Number 9004000
Medical Device Problem Codes Inappropriate Waveform (2536); Intermittent Loss of Power (4016)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was received for evaluation.The rear sample line was damaged.Flowrate test and hi/lo calibration testing were done and the reported problem was duplicated.The stated problem of: not giving proper wave form/ pump starts and stops/ eval ac adapter, was verified.For the wave form problem and pump starts and stops, the pump flow was slow and the pump was very loud and grinding.The pump was replaced and the monitor passed the hi/lo calibration and the wave form was steady with a flowrate within specification.As for the evaluation on the ac cord, the strain relief was damaged and had a steady 24 volts.The problem was caused by a defective co2 pump.Malfunctioning co2 pump mechanism no longer drawing adequate airflow from wear of age.Due to wear of life the co2 pump has reached the end of its useful life after 6 years of operation.The device history record (dhr) review is not applicable or relevant because the results of the complaint investigation do not indicate a problem with the initial manufacture or prior repair of the device.This was the first time the device was received for service after 6 years in the field and the failure resulted from worn co2 pump mechanism due to wear of age.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
 
Event or Problem Description
It was reported that the device was not giving the proper wave form, the pump starts and stops, and the ac adaptor was worn.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BCI CAPNOCHECK PLUS
Common Device Name
ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15422665
Report Number3012307300-2022-19201
Device Sequence Number4749987
Product Code CCK
UDI-Device Identifier10610586036811
UDI-Public10610586036811
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
K970209
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 09/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number9004000
Device Catalogue Number9004000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/17/2020
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/23/2022
Initial Report FDA Received Date09/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-