Model Number 75017089
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/25/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that the polarcup head/liner inserter blue head is broken.It is unknown whether this problem was noticed in a surgical environment or not; therefore, patient involvement has not been confirmed.
 
Additional Manufacturer Narrative
H3, h6: it was reported that the polarcup head/liner inserter blue head is broken.It is unknown whether this problem was noticed in a surgical environment or not; therefore, patient involvement has not been confirmed.The device in scope of this complaint, was returned for investigation.The reported issue could be confirmed upon visual inspection.One plastic part was found fractured.A review of the batch record revealed no deviations from the standard manufacturing process that could have contributed to the reported issue.The complaint history review was performed for the affected batch f71731.A number of 7 further complaints were found for devices of the same lot with the same failure mode.However, a batch size of 1000 pieces was produced.The additional complaints are not determined to indicate a manufacturing related issue with this batch.The occurrence of the reported failure mode is within its expected risk level as per risk management.A review of the risk management documentation verified the failure mode and severity of the reported issue.A review of past corrective actions was performed.No further escalation is required.There are no indications that the complaint device did not meet manufacturing specifications upon release for distribution.Based on the performed investigations, the reported failure mode could be confirmed.A relationship between the reported event and the device can be confirmed.No probable cause can be determined.Normal wear and tear through repeated use is known to contribute to the reported event.According to document "processing (cleaning, disinfection and sterilization) of instruments from smith & nephew orthopaedics ag" (lit.N°03389-en 1363 v3 11/19), all devices must be inspected and controlled for proper functioning after cleaning/disinfection.No further actions are deemed necessary at this time.This version of the device will be monitored for similar issues.This investigation is considered closed.The complaint device will be discarded.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLARCUP HEAD/LINER INSERTER
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS AG
schachenallee 29
aarau CH-50 00
SZ  CH-5000
Manufacturer (Section G)
SMITH & NEPHEW ORTHOPAEDICS AG
schachenallee 29
aarau CH-50 00
SZ   CH-5000
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15426445
Report Number9613369-2022-00509
Device Sequence Number7502123
Product Code LXH
UDI-Device Identifier07611996103151
UDI-Public07611996103151
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K110135
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number75017089
Device Catalogue Number75017089
Device Lot NumberF71731
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/07/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/25/2022
Supplement Date Received by Manufacturer10/10/2022
Initial Report FDA Received Date09/15/2022
Supplement Report FDA Received Date10/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/12/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-