The device was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, per complaint details, a revision was performed due to a broken post after an unknown length of time in-vivo.As of the date of this medical investigation, no supporting clinical documentation has been provided; therefore, no clinical factors could be assessed which would have contributed to the event.With the limited information provided, the patient impact beyond the reported post breakage and revision cannot be determined.No further medical assessment can be rendered at this time.A review of complaint history of the previous 12 months revealed similar events for the listed device, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the instructions for use documents for knee systems revealed that the implant can break or become damaged as a result of strenuous activity or trauma, and has a finite expected service life and may need to be replaced in the future.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.Verifying the device configuration according to the drawing print is within the steps of the final inspection, additionally, control of the quality and manufacture of 7.5 mrad cross-linked ultra-high-molecular-weight polyethylene (uhmwpe) bar stock is part of the material specifications.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Factors that could contribute to the reported event include traumatic injury, size selected, surgical technique or postoperative care.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|