(b)(6).Reporter occupation = sales rep.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Event summary: as reported, a hair was found inside the sterile packaging of a guardia accesset embryo transfer catheter.The device did not make patient contact.Another device of the same type was used to complete the procedure.There were no adverse effects reported.Investigation - evaluation: reviews of the complaint history, device history record, instructions for use, specifications, and quality control procedures and a visual inspection of the device were conducted during the investigation.The product was returned in sealed packaging.Foreign matter was visible inside of the sealed packaging.The outer package was opened to confirm location.The foreign matter was within the sealed packaging of the guide catheter.A document-based investigation evaluation was performed.No related non-conformances were recorded, and there have been no other reported complaints for this lot number.There is no evidence of additional nonconforming devices from the complaint lot in house or in the field.Sufficient inspection activities are in place to identify this failure mode prior to distribution.The device is provided with instructions for use which state, "supplied sterilized by ethylene oxide gas in peel-open packages.Intended for one-time use.Sterile if package is unopened and undamaged.Do not use the product if there is doubt as to whether the product is sterile.Store in a dark, dry, cool place.Avoid extended exposure to light.Upon removal from the package, inspect the product to ensure no damage has occurred.¿ based on the available information, cook has concluded that it is most likely the packaging quality control inspector missed the foreign matter when inspecting entire packaged product and seal for particulate.The operator involved completed awareness training for this nonconforming product.Cook will continue monitoring of similar complaints and has notified the appropriate personnel of this event.Per a review of risk documentation, no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|