Model Number LPG1510 |
Device Problems
Loss of or Failure to Bond (1068); Product Quality Problem (1506); Material Split, Cut or Torn (4008)
|
Patient Problems
Hernia (2240); Unspecified Tissue Injury (4559)
|
Event Date 08/24/2022 |
Event Type
Injury
|
Event Description
|
According to the reporter, on post operative bilateral robotic inguinal hernia repair, the patient came back as he was still having symptoms of feeling a bulge.The surgeon took a look laparoscopically and noticed that the mesh was damaged.There was a small hole in the center of the mesh.The surgeon did not think this was causing the symptoms but also noticed that there was a slight indentation far medially that the mesh did not cover and the patient recurred.The surgeon had to re-repaired it via open procedure and the fixation system that was used to close the peritoneum was suturing.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was not returned, but a photo was available for evaluation.Visual inspection noted a hole was found.It was reported that the mesh tore, the mesh was difficult to adhere, did not stay in place, or shifted within the patient while being implanted, and the physical appearance of the product was different than expected.The reported issues were confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: grasp the mesh at either end and insert it through the trocar.It is recommended to use a trocar of at least 10mm internal diameter to introduce a mesh of size up to 15x10 cm and a trocar of at least 12mm internal diameter to introduce a mesh of size 16x12 cm or above.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
Search Alerts/Recalls
|
|
|