| Model Number |
UH400U |
| Medical Device Problem Codes |
Output Problem (3005); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/16/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned for evaluation; the evaluation has not yet been completed.
|
| |
|
Event or Problem Description
|
|
Customer reported that the autocon iii unit stopped working during a procedure and the case had to be canceled as they did not have a backup unit.There was no injury to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Per the evaluation findings: faulty control board.Easy monitor 300 ohm test 285-315 ohm, reading 330 ohm.Primary cause: wear/often used.The device had been in use for approximately 3 years.
|
| |
|
Search Alerts/Recalls
|