MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON CLAREON ASPHERIC UV ABSORBING IOL; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON CLAREON ASPHERIC UV ABSORBING IOL; INTRAOCULAR LENS Back to Search Results
Model Number CC60WF
Medical Device Problem Code Scratched Material (3020)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/10/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A facility representative reported that during an intraocular lens (iol) implant procedure, it was found that iol was scratched.The iol replaced with another iol during initial procedure additional information received and stated that the patient was not hospitalized for the event and there was no patient harm.
 
Additional Manufacturer Narrative
The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
Additional information received.And stated, that no suture was added.
 
Additional Manufacturer Narrative
The lens and the associated company cartridge were returned inside a plastic specimen cup with screw top lid.Only two cut portions of the lens were returned.Solution was observed dried on the lens pieces.The optic was scratched on the anterior and the posterior near the center.The cut portions had additional cracked areas.The optic edge has a nicked area.The used non-qualified company cartridge has an inadequate amount of viscoelastic observed.The cartridge tip has stress lines.The cartridge wings display evidence of handpiece use.The account indicated the use of a non-qualified cartridge.The diopter of this company lens (26.0 diopter) is beyond the range that has been qualified for use in a company cartridge.A handpiece and viscoelastic were indicated that are qualified when used with this lens and a qualified company cartridge combination.The reported scratch damage was observed.The root cause is most likely related to a failure to follow the ifu (instructions for use).The company lens has been qualified for diopters 6.0 through 25.0 with a company, and diopters 25.5 through 30.0 are qualified for use in the company cartridge.Per the ifu: the company iol delivery system is for implantation of qualified company foldable iols.No unqualified lenses should be used with the company iol delivery system.The company cartridges are qualified for use with compatible company handpieces for the surgical implantation of company qualified foldable iols.Company foldable iols are qualified for use with an company qualified delivery system (handpiece and cartridge) and ophthalmic viscosurgical device (ovd) combination.The use of an unqualified combination may cause damage to the iol and potential complications during the implantation process.Also an inadequate amount of viscoelastic was observed in the cartridge.The ifu instructs to completely fill the cartridge with ovd (diagram provided) immediately prior to loading and delivery of the lens.Do not attempt to load the lens without adequate ovd in the device.Not adequately filling the device with viscoelastic will result in inadequate coverage of lens and the lens fold path with ovd, which may result in damage.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLAREON ASPHERIC UV ABSORBING IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key15427672
Report Number1119421-2022-01963
Device Sequence Number4750140
Product Code HQL
UDI-Device Identifier00380652396943
UDI-Public00380652396943
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
P190018
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/17/2025
Device Model NumberCC60WF
Device Lot Number15308090
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 08/23/2022
Supplement Date Received by Manufacturer10/31/2022
11/30/2022
Initial Report FDA Received Date09/15/2022
Supplement Report FDA Received Date11/28/2022
12/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/19/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CLAREON MONARCH IV IOL DELIVERY SYSTEM, INJECTOR; DUOVISC; MONARCH III CARTRIDGE D; PROVISC
Patient Age72 YR
Patient SexFemale
Patient Weight173 KG
-
-