• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT POINT OF CARE I-STAT CREA CARTRIDGE; CREA+ CARTRIDGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT POINT OF CARE I-STAT CREA CARTRIDGE; CREA+ CARTRIDGE Back to Search Results
Catalog Number 03P84-25
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code Abdominal Pain (1685)
Date of Event 09/08/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Apoc incident # (b)(4).Apoc labeling will be evaluated during the investigation as pertaining to the event.
 
Event or Problem Description
On (b)(6) 2022, abbott point of care was contacted by a customer regarding i-stat crea cartridges that yielded suspected discrepant creatinine result of <0.2 mg/dl on a 50 year old male patient with abdominal pain.There was no additional patient information available at the time of this report.Return product is available for investigation.(b)(6).There are no injuries associated with this event.At this time there is no reason to suspect a malfunction exists.The customer also reports that the patient was given contrast based on one i-stat result of <0.2 mg/dl.The investigation is underway.
 
Additional Manufacturer Narrative
Apoc incident: # (b)(4).The investigation was completed on (b)(6) 2022.A review of the device history record confirmed the cartridge lot passed finished goods release criteria.Retained cartridge testing met the acceptance criteria found in q04.01.003 rev.Ak, appendix 1- product complaint level 2 and level 3 investigation procedure.Returned cartridge testing met the acceptance criterion for suppressed results but the sample size was too low (<17) to assess the rate of points outside total allowable error (ea).No deficiency has been identified.
 
Event or Problem Description
Na.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
I-STAT CREA CARTRIDGE
Common Device Name
CREA+ CARTRIDGE
Manufacturer (Section D)
ABBOTT POINT OF CARE
400 college road
princeton NJ 08540 6607
Manufacturer (Section G)
ABBOTT POINT OF CARE CANADA LTD.
185 corkstown road
K2H 8 V4
CA   K2H 8V4
Manufacturer Contact
linda maczuszenko
400 college road
princeton, NJ 08540
6136885949
MDR Report Key15427829
Report Number2245578-2022-00136
Device Sequence Number15380714
Product Code CGL
UDI-Device Identifier10054749000125
UDI-Public10054749000125
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
K973292
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/28/2022
Device Catalogue Number03P84-25
Device Lot NumberA22182A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/09/2022
Supplement Date Received by Manufacturer09/28/2022
Initial Report FDA Received Date09/15/2022
Supplement Report FDA Received Date10/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/01/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age50 YR
Patient SexMale
-
-