• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION CLEARLINK CONTINU-FLO STOPCOCK AND I.V. SOLUTION ADMINISTRATION SETS; STOPCOCK, I.V. SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION CLEARLINK CONTINU-FLO STOPCOCK AND I.V. SOLUTION ADMINISTRATION SETS; STOPCOCK, I.V. SET Back to Search Results
Catalog Number 2C6255
Medical Device Problem Codes Disconnection (1171); Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
These events occurred (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was reported that the tubing of an unspecified quantity of clearlink system continu-flo solution set large bore stopcock extension sets disconnected and leaked.The event was further described as; ¿iv (intravenous) tubing coming apart at some of the connections or spikes that are falling out of the bag¿ resulting in leaks.It was not specified when in the process step this occurred.There was no report of patient injury or medical intervention associated with these events.No additional information is available.
 
Additional Manufacturer Narrative
H10: a batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The devices were discarded; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEARLINK CONTINU-FLO STOPCOCK AND I.V. SOLUTION ADMINISTRATION SETS
Common Device Name
STOPCOCK, I.V. SET
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - CARTAGO
see h10
see h10
cartago 30106
CS   30106
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key15428433
Report Number1416980-2022-04845
Device Sequence Number17313299
Product Code FMG
UDI-Device Identifier00085412046310
UDI-Public(01)00085412046310
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K130245
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number2C6255
Device Lot NumberR22A13116
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/18/2022
Supplement Date Received by Manufacturer10/13/2022
Initial Report FDA Received Date09/15/2022
Supplement Report FDA Received Date10/14/2022
Date Device Manufactured01/15/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-