• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR GRAFTMASTER RX CORONARY STENT GRAFT SYSTEM; CORONARY STENT DELIVERY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR GRAFTMASTER RX CORONARY STENT GRAFT SYSTEM; CORONARY STENT DELIVERY SYSTEM Back to Search Results
Model Number 1012580-26
Medical Device Problem Codes Difficult to Remove (1528); Difficult to Advance (2920)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/22/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the procedure was to treat a highly calcified right coronary artery (rca).An atherectomy device and was able to cross the lesion after four attempts.Angiography revealed a perforation in the proximal mid vessel.A 2.5x20mm non-abbott balloon was inflated and a second access was obtained in the left common femoral artery.Due to patient's debility, pre-dilatation was performed with a 3.0x15 non-abbott balloon at 12 atmospheres (atms) at 20 seconds with multiple inflations throughout the proximal mid vessel.A 2.8x16mm graftmaster stent delivery system (sds) was attempted to cross into the rca; however, it failed as the stent became stuck with at the proximal portion of the guide catheter.Delay to treat the perforation was noted due to the failure to cross.Another 2.8x26mm graftmaster sds was attempted to cross through the left in the rca; however, also became stuck with the guide catheter.Delay to treat the perforation was noted due to the failure to cross.A 2.75x38mm non-abbott stent was deployed mid vessel, but perforation was still noted.A 3.0x18mm non-abbott balloon was inflated to 15 atms 20 seconds throughout the length of the stent; however, the perforation was still noted.Therefore, another 3.5x18mm non-abbott balloon was inflated to 15 atms, but perforation was still present.The patient was hemodynamically stable throughout this.A 4.0x26mm graftmaster covered stent was deployed at 15 atms for 20 seconds and the stent was apposed, but not well expanded.A 4.0x18mm non-abbott balloon was used to post-dilate; however, perforation was still present.After attempts to cross a 4.0x16mm graftmaster stent to the proximal edge, the shaft of the stent broke and the balloon ruptured without deploying.This was removed; however, unknown how at this moment.A 4.0x30mm non-abbott balloon was advanced distally, inflated at 15 atms and no dissection, perforation, distal embolization or side branch loss noted.The patient became hypoxic and there was not significant pulse.Cardiopulmonary resuscitation (cpr) was performed for 5 to 10 minutes.Percutaneous coronary intervention in the ostial proximal to mid rca with a 2.75x38mm non-abbott stent and followed by a 4.0x16mm non-abbott stent.Thrombus was noted in the mid vessel and was treated with a non-abbott balloon.An impella cp placement was performed with a.035 supra core wire in the left ventricular cavity.During patient's initial episode at times towards the end the patient was episodically hypotensive and a pericardiocentesis needle was used to puncture the pericardium.Echocardiogram revealed an increase effusion.Pericardiocentesis was performed and a catheter was left in place for a pericardial drain.Patient was stabilized and perforation had sealed.The patient was take to cardiovascular icu in critical state.The patient expired later on the same day.No additional information was provided.
 
Additional Manufacturer Narrative
The device is returning for analysis; however, it has not been received.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.The three additional graftmaster stents referenced are filed under separate medwatch report numbers.
 
Additional Manufacturer Narrative
A visual and dimensional inspection was performed on the returned device.The reported difficult to advance and difficult to remove could not be tested due to the condition of the returned device.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history did not indicate a lot specific quality issue.A conclusive cause for the reported difficulties could not be determined.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
 
Event or Problem Description
Subsequently, after the initial report was filed, the account confirmed the 4.0x16mm graftmaster separated into two pieces at the shaft and the separated portion was removed.The account confirmed there was no intervention to remove the separated portion it was simply pulled out.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GRAFTMASTER RX CORONARY STENT GRAFT SYSTEM
Common Device Name
CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key15428675
Report Number2024168-2022-09745
Device Sequence Number7708945
Product Code MAF
UDI-Device Identifier08717648176364
UDI-Public08717648176364
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
H000001
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2023
Device Model Number1012580-26
Device Catalogue Number1012580-26
Device Lot Number1052541
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 08/22/2022
Supplement Date Received by Manufacturer10/31/2022
Initial Report FDA Received Date09/15/2022
Supplement Report FDA Received Date11/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/25/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
.014 PROWATER GUIDE WIRE; GRAND SLAM WIRE; TWIN PASS CATHETER
Outcome Attributed to Adverse Event Other;
Patient Age84 YR
Patient SexMale
-
-