| Model Number |
L98110 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
01/01/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Hcp is reporting a distal fracture of corail stem.Doi: (b)(6) 2022; dor: (b)(6) 2022; unknown hip.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary : the device associated with this report was received for analysis.Based in the visual examination of the part, it was confirmed that stem was fractured at the distal portion of the device's plug.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.Device history lot : bfarm dhr review request: product description: corail revision stem ho 10, product code: 5262301, lot number: l98110.Please provide: 1) quantity manufactured: (b)(4), 2) date of manufacture: 03/16/2016, 3) any anomalies or deviations identified in dhr: none.No deviation or non-conformance was identified.4) expiry date: 02/28/2021, 5) ifu reference: w90956 rev.C.Device history review : a manufacturing record evaluation was performed for the finished device (5262301/l98110) product and lot numbers, and no non-conformances were identified.
|
| |
|
Search Alerts/Recalls
|