It was reported the safe-t-j fixed core wire guide malfunctioned.While using the wire guide to place the drain, the wire separated in the patient.At this time, no other adverse effects have been reported for this incident.Additional information regarding the event and patient outcome has been requested but is currently unavailable.
|
|
Investigation ¿ evaluation: on 09sep2022, cook medical inc.Received a complaint from a representative at the baptist memorial hospital, located in the city of southaven, ms.It was reported the safe-t-j fixed core wire guide (rpn:tscf-35-80-3; lot number 14903820) unraveled while in the patient when placing a drain.Reviews of documentation including the complaint history, device history record (dhr), manufacturing instructions (mi), quality control procedures, and specifications of the complaint device were conducted during the investigation.The complaint device was not returned; therefore, no physical examinations could be performed.However, the customer did provide a photo of the reported difficulty.A visual examination confirmed a small section near the distal end of the wire guide to be unravelling resulting in the coil becoming elongated.Additionally, a document-based investigation evaluation was performed.A review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.A review of the dhr for the reported complaint device lot 14903820 recorded no related non-conformances.To date, a further search of our database records revealed no additional complaints associated with the reported lot.Since there is objective evidence the dhr was fully executed and there are no other lot-related complaints from the field, cook has concluded that there is no evidence that non-conforming product exists in house or in the field, and that the device was manufactured to current specifications.The tscf-35-80-3, safe-t-j fixed core wire guide is not supplied with an instructions for use (ifu) booklet.Based on the information provided, no product returned, and the results of the investigation, cook determined the root cause to be component failure without any design or manufacturing issue.The appropriate personnel have been notified.Cook will continue to monitor for similar complaints.Per the risk assessment no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|