|
ETHICON INC. PRLNE BLU 30IN 5-0 S/A C-1; SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE
|
Back to Search Results |
|
| Model Number |
8890H |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/08/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Events reported via: 2210968-2022-07584, 2210968-2022-07585, 2210968-2022-07586, 2210968-2022-07588, 2210968-2022-07589, 2210968-2022-07591.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a pancreas procedure on (b)(6) 2022 and suture was used.During the procedure, the suture broke when the surgeon tied the first knot when suturing a vessel.The tried with 6 additional sutures from the same lot and all of them broke when he tied the first knot.The or time was prolonged due to this.A different device was used to complete the procedure.No adverse patient consequences were reported.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Date sent to the fda: 11/7/2022.H6 component code: g07002 no device problem found.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.H3 investigational summary: the product was returned to ethicon for evaluation.Visual inspection and functional test evaluation were conducted on the returned device.Visual analysis of the returned sample determined that twenty packets that pertains to the product code 8890h were returned to ethicon for analysis with the packaging closed.Upon initial inspection, of the samples, no external damages were observed on the packets.As per the sampling plan, a visual inspection was performed on thirteen samples, and no defects were found on the packages.The samples were opened, and the swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along the strand and no defects, damage, or suture breakage were noted.A functional test was performed using an instron equipment and the tensile strength force was above the minimum requirement.The event described could not be confirmed as the device performed without any defect noted.It should be noted that as part of our quality process, each batch is randomly inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Related reports: 2210968-2022-07584, 2210968-2022-07585, 2210968-2022-07586, 2210968-2022-07588, 2210968-2022-07589, 2210968-2022-07591.
|
| |
|
Search Alerts/Recalls
|
|
|