Model Number H7493941828350
Medical Device Problem Code Break (1069)
Health Effect - Clinical Codes Hemorrhage/Blood Loss/Bleeding (1888); Perforation (2001); Foreign Body In Patient (2687)
Date of Event 08/22/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Age at time of event - 18 years or older.
 
Event or Problem Description
It was reported that a shaft break occurred; the patient experienced vessel rupture with bleeding, resulting in surgery.The target lesion was located in the left anterior descending artery (lad).The patient had angiograms performed, and intervention was attempted on the left anterior descending artery.A 3.50 x 28mm synergy xd drug-eluting stent (des) was advanced for treatment, however, during stent inflation, the left anterior descending artery ruptured with bleeding into the myocardium.Upon removal of the system it was discovered the balloon broke with fragments in the lad.An emergent sternotomy/ligation was performed to repair the lad.Ultimately the patients lad was not able to resolve and rapid, massive blood infusion was needed.No bypass was performed as the patient has large circumflex and right coronary artery.The patient was then transferred to cardiovascular intensive care unit.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY XD
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15429367
Report Number2124215-2022-33483
Device Sequence Number17325842
Product Code NIQ
UDI-Device Identifier08714729981077
UDI-Public08714729981077
Combination Product (Y/N)Y
Initial Reporter StateMI
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/15/2024
Device Model NumberH7493941828350
Device Lot Number0028931007
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/25/2022
Initial Report FDA Received Date09/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/15/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age64 YR
Patient SexMale
-
-