| Model Number |
1012405-20 |
| Medical Device Problem Code |
Difficult to Remove (1528)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/25/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was received.The investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.The bmw universal guide wire referenced in b5 and d10 is filed under a separate medwatch report number.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was performed to treat a heavily calcified lesion in the mid left anterior descending artery.A roto blade catheter was advanced and rotablation was performed on the artery.A 2.5x20mm trek otw balloon dilatation catheter (bdc) was backloaded onto the roto blade wire and inflated once to 8 atmospheres (atms).The bdc was deflated without issues and the roto blade wire removed.A hi-torque bmw universal guide wire was advanced through the bdc and the bdc was inflated once again to 8 atms.The bdc was deflated without issues; however, when attempting to remove the bdc from the guide wire it was stuck.Both the bdc and the guide wire had to be removed as a unit.The procedure was successfully completed with a new bmw and trek bdc.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual inspection was performed on the returned device.The reported difficulty removing the device from the guide wire was confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation determined the reported complaint appears to be related to circumstances of the procedure.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.Na.
|
| |
|
Search Alerts/Recalls
|