• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION POLARIS X; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION POLARIS X; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING Back to Search Results
Model Number 86080
Medical Device Problem Codes Material Perforation (2205); Defective Device (2588)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/18/2022
Type of Reportable Event Malfunction
Event or Problem Description
During preparation for a procedure as two polaris x catheters were taken out of the box, there were scratches on the catheter shafts where the blue plastic was pulling away.The procedure was completed with a new device and without any patient complications.The catheters have been returned for analysis.
 
Event or Problem Description
During a procedure a polaris x catheter was selected for use.There were scratches on the catheter shaft where the blue plastic was pulling away.The scratches were observed upon opening the catheter out of the box.The procedure was completed with a new device and without any patient complications.
 
Additional Manufacturer Narrative
The device was returned to boston scientific for analysis.Upon analysis visual inspection revealed residues of adhesive were found around the shaft.Therefore, the reported complaint is confirmed.Correction: e1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLARIS X
Common Device Name
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
302 parkway, global park
la aurora - heredia
CS  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
dc a330
saint paul, MN 55112
6515826168
MDR Report Key15429571
Report Number2124215-2022-36180
Device Sequence Number4622328
Product Code DRF
UDI-Device Identifier08714729270188
UDI-Public08714729270188
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K003452
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 12/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/01/2024
Device Model Number86080
Device Catalogue Number86080
Device Lot Number0029323483
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/09/2022
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/18/2022
Supplement Date Received by Manufacturer11/08/2022
Initial Report FDA Received Date09/15/2022
Supplement Report FDA Received Date12/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/02/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-