• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. DISPOSABLE CUFF WITH PLC AND PROTECTIVE SLEEVE, 18 IN. (46CM) - DP, SB; TOURNIQUET CUFF, SINGLE-USE; TOURNIQUET CUFF BANDAGE, NON-LATEX

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. DISPOSABLE CUFF WITH PLC AND PROTECTIVE SLEEVE, 18 IN. (46CM) - DP, SB; TOURNIQUET CUFF, SINGLE-USE; TOURNIQUET CUFF BANDAGE, NON-LATEX Back to Search Results
Catalog Number 60707015300
Medical Device Problem Code Gas/Air Leak (2946)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/24/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This complaint is recorded by zimmer biomet under (b)(4).A follow up/ final report will be submitted once investigation is complete.
 
Event or Problem Description
It was reported that the cuff had air leakage.The event timing was during preparation of surgery.There is no harm or delay reported.No adverse events were reported as a result of this malfunction.
 
Additional Manufacturer Narrative
This complaint is recorded by zimmer biomet under (b)(4).When connected to a known working ats in the lab, the ats had to keep pumping air to keep the cuff inflated, confirming a leak.A bubble test determined the leak was located where the pvc tubing connects to the right-angle connectors of the cuff.Review of the device history record identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
There is no additional event information available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DISPOSABLE CUFF WITH PLC AND PROTECTIVE SLEEVE, 18 IN. (46CM) - DP, SB
Common Device Name
TOURNIQUET CUFF, SINGLE-USE; TOURNIQUET CUFF BANDAGE, NON-LATEX
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
5745260384
MDR Report Key15430934
Report Number0001526350-2022-00870
Device Sequence Number7511367
Product Code KCY
UDI-Device Identifier00889024377172
UDI-Public(01)00889024377172(17)170206(10)Z000000000
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 12/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/28/2025
Device Catalogue Number60707015300
Device Lot Number45815358
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/01/2022
Supplement Date Received by Manufacturer12/15/2022
Initial Report FDA Received Date09/15/2022
Supplement Report FDA Received Date12/16/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/29/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Patient SexUnknown
-
-