| Model Number |
37262E |
| Medical Device Problem Code |
Component Missing (2306)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/22/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported while using bd alaris smartsite extension set the end cap was missing.There was no report of patient impact.The following information was provided by the initial reporter: extension tubing noted to be missing an end cap.
|
| |
|
Additional Manufacturer Narrative
|
|
Date of event is unknown; awareness date has been used for this field.Medical device expiration date: unknown.Device manufacture date: unknown.Address information was not able to be obtained, therefore, nj was used as a place holder.A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported while using bd alaris smartsite extension set the end cap was missing.There was no report of patient impact.The following information was provided by the initial reporter: extension tubing noted to be missing an end cap.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: investigation summary a complaint of a set missing an end cap was received from the customer.Photos of the material and lot were provided by the customer.There was also a photo that showed the smart site missing a component.The customer complaint was confirmed.A device history record review for model 37262e lot number 21126091 was performed.The search showed that a total of 8,803 units in 1 lot number were built on 22dec2021.There were no quality notifications issued for the failure mode reported by the customer during the production build of this set.As a physical sample was not returned, a root cause could not be determined.
|
| |
|