| Model Number |
C0R83 |
| Medical Device Problem Code |
Device Contaminated During Manufacture or Shipping (2969)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/25/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The event device is anticipated to return.A follow-up report will be provided upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
Name of procedure being performed: na (incoming inspection).Detailed description of event: [facility] incoming inspection po#: (b)(4).This is a complaint from [facility] evaluation team.The lot mentioned above was found to have defective units per our incoming inspection.Delivery date: july 22, 2022.Please refer to the attachment file.We found defects as the result of incoming inspection of the products.Please refer to the attachment material for the quantity which returns the products.We found the defects that already informed applied medical by cer.We just inform you the defect and return the products for following defects.You do not need investigation.Hairlike object in pouch - c0r83 lot#: 1425082 (1 ea.).Patient status: (no patient involvement).Type of intervention: (incoming inspection).
|
| |
|
Event or Problem Description
|
|
Name of procedure being performed: na (incoming inspection).Detailed description of event: [facility] incoming inspection po# (b)(4).This is a complaint from [facility] evaluation team.The lot mentioned above was found to have defective units per our incoming inspection.Delivery date: july 22, 2022.Please refer to the attachment file.We found defects as the result of incoming inspection of the products.Please refer to the attachment material for the quantity which returns the products.We found the defects that already informed applied medical by cer.We just inform you the defect and return the products for following defects.You do not need investigation.Hairlike object in pouch - c0r83 lot #1425082 (1 ea.) patient status: na (no patient involvement).Type of intervention: na (incoming inspection).
|
| |
|
Additional Manufacturer Narrative
|
|
The event unit was returned to applied medical for evaluation.Visual inspection confirmed the presence of hair inside the sterile unopened pouch.Based on the condition of the returned unit and the description of the event, the hair likely originated from an operator during the manufacturing process.
|
| |
|
Search Alerts/Recalls
|