| Model Number |
9300P-1001 |
| Medical Device Problem Code |
Unable to Obtain Readings (1516)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Complainant alleged that during functional testing, the device was unable to obtain an ecg signal via pads.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation for evaluation.The customer's report was observed during review of the device activity log.However, the device was put through extensive testing including functionality testing without duplicating the report.Internal inspection of the device found no discrepancies.No accessories were returned as part of this investigation.The main board was replaced and scrapped as a precaution.The device was recertified and returned to the customer.Analysis for reports of this type has not identified an increase in trend.
|
| |
|
Search Alerts/Recalls
|