• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ULTRAFLEX ESOPHAGEAL NG; PROSTHESIS, ESOPHAGEAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ULTRAFLEX ESOPHAGEAL NG; PROSTHESIS, ESOPHAGEAL Back to Search Results
Model Number M00513850
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/22/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Patient's exact age is unknown; however it was reported that the patient was over the age of 18.The complainant was unable to provide the suspect device lot number.Therefore, the manufacture and expiration dates are unknown.Imdrf device code a0413 captures the reportable event of silastic tube on the ultraflex device fell inside the patient and was removed with forceps.
 
Event or Problem Description
It was reported to boston scientific corporation on august 22, 2022, that an ultraflex esophageal ng proximal release covered stent was implanted to treat a stricture due to esophageal cancer during an esophageal stent placement procedure performed on (b)(6), 2022.During the procedure, the ultraflex esophageal stent was successfully deployed.However, it was noted that the silastic band on the delivery system fell inside the patient.The silastic band was removed using forceps.The stent remained implanted and the procedure was completed with this device.There were no patient complications reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRAFLEX ESOPHAGEAL NG
Common Device Name
PROSTHESIS, ESOPHAGEAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key15433028
Report Number3005099803-2022-05356
Device Sequence Number20334826
Product Code ESW
UDI-Device Identifier08714729716075
UDI-Public08714729716075
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K091816
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberM00513850
Device Catalogue Number1385
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/22/2022
Initial Report FDA Received Date09/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-