The treatment tip was returned and the evaluation was completed.During evaluation of the treatment tip, service found dielectric breakdown.The tip passed the flow test and thermistor test and failed the leak test.The tip failed the visual inspection as dielectric breakdown was observed.No functional testing was performed due to the dielectric breakdown.According to the thermage cpt system technical user¿s manual, burns are a known possible reaction to treatment.The procedure may produce heating in the upper layers of the skin, causing burns and subsequent blister and scab formation.A review of the manufacturing records showed all requirements were met.The lot history, trend analysis, risk analysis and/or directions for use review were considered acceptable, with the product performing within anticipated rates.Final test verification specifications are acceptable.No non-conformities or anomalies found related to this complaint when reviewing the device history record.Based on the available information, this event was most likely caused by damage on the tip membrane.No corrective action is required.
|