• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON & CO. (BROKEN BOW) BD VACUTAINER® BUFFERED SODIUM CITRATE (9NC) BLOOD COLLECTION TUBES; BLOOD SPECIMEN COLLECTION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON & CO. (BROKEN BOW) BD VACUTAINER® BUFFERED SODIUM CITRATE (9NC) BLOOD COLLECTION TUBES; BLOOD SPECIMEN COLLECTION DEVICE Back to Search Results
Model Number 369714
Medical Device Problem Code Short Fill (1575)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/30/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported when using the bd vacutainer® buffered sodium citrate (9nc) blood collection tubes experience underfilling tubes when they are about to expire.The following information was provided by the initial reporter.The customer stated: "it was reported by the customer that tubes are underfilling and it is "normal" for them to experience underfilling tubes when they are about to expire.".
 
Additional Manufacturer Narrative
The following fields were updated due to additional information: d9: device available for eval: yes.D9: returned to manufacturer on: 2022-09-26.H.6.Investigation summary: bd received 1 sample from lot 1350276 and 5 samples from lot 1350273in support of this complaint.The sample tubes were functionally tested and the issue of underfill was not observed.Additionally, 10 retention tubes from the bd inventory were functionally tested and the issue of underfill was not observed.Bd was unable to confirm and/or duplicate the customer¿s indicated failure mode because the defect was not observed in the sample and retention testing.The device history records for both lots were reviewed with no issues being identified.There were no related quality notifications.All process and final inspections comply with specification requirements.H3 other text : see h.10.
 
Event or Problem Description
It was reported when using the bd vacutainer® buffered sodium citrate (9nc) blood collection tubes experience underfilling tubes when they are about to expire.The following information was provided by the initial reporter.The customer stated: "it was reported by the customer that tubes are underfilling and it is "normal" for them to experience underfilling tubes when they are about to expire.".
 
Event or Problem Description
It was reported when using the bd vacutainer® buffered sodium citrate (9nc) blood collection tubes experience underfilling tubes when they are about to expire.The following information was provided by the initial reporter.The customer stated: "it was reported by the customer that tubes are underfilling and it is "normal" for them to experience underfilling tubes when they are about to expire.".
 
Additional Manufacturer Narrative
The following fields have been updated with additional information: there were multiple lot numbers reported to be involved.The information for each lot number is as follows: d.4.Medical material #: 363083.D.4.Medical device lot #: 1350273.D.4.Medical device expiration date: 2022-09-30.H.4.Device manufacture date: 2021-12-16.D.4.Medical device lot #: 1350276.D.4.Medical device expiration date: 2022-09-30.H.4.Device manufacture date: 2021-12-16.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® BUFFERED SODIUM CITRATE (9NC) BLOOD COLLECTION TUBES
Common Device Name
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BECTON, DICKINSON & CO. (BROKEN BOW)
150 south 1st avenue
broken bow NE 68822
Manufacturer (Section G)
BECTON, DICKINSON & CO. (BROKEN BOW)
150 south 1st avenue
broken bow NE 68822
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15433384
Report Number1917413-2022-00572
Device Sequence Number5594915
Product Code JKA
UDI-Device Identifier50382903697149
UDI-Public50382903697149
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K945952
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number369714
Device Catalogue Number363083
Device Lot NumberSEE H.10
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/02/2022
Supplement Date Received by Manufacturer09/29/2022
10/10/2022
Initial Report FDA Received Date09/15/2022
Supplement Report FDA Received Date10/05/2022
10/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-