• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH 32MM +8 V40 TAPER VIT HEAD; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH 32MM +8 V40 TAPER VIT HEAD; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Model Number 6260-5-332
Medical Device Problem Codes Degraded (1153); Fracture (1260); Detachment of Device or Device Component (2907); Unintended Movement (3026)
Health Effect - Clinical Code Foreign Body Reaction (1868)
Date of Event 08/22/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.Device not returned to the manufacturer.
 
Event or Problem Description
As reported by rep: "case report: patient who underwent surgery approximately 12 years ago for a left total hip arthroplasty, presented with pain and x-rays showed a femoral head disassociation from the trunnion of the stem.The original stryker implants implanted were: accolade tmzf size 3x132 stem, a 32+8 v40 cocr femoral head, a size 50 trident psl ha hemispherical acetabular shell, with no screws, and size 32 e 10° x3 polyethylene liner.The patient was brought into the operating room, a revision hip was performed on the left side, the head was fractured from the trunnion on the stem.Rep confirmed the head had worn the trunnion down and became dissociated, the stem did not fracture, the head was free in the acetabulum causing impingement and metallosis.There was adverse local tissue reactions from the trunnionosis however the adductors were intact.Another company's revision stem was used, the trident cup was retained and a new 36mm size e 10 0° liner was put into the cup.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
32MM +8 V40 TAPER VIT HEAD
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-CORK
ida industrial estate
carrigtwohill NA
EI   NA
Manufacturer Contact
arokiya raj
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key15433456
Report Number0002249697-2022-01352
Device Sequence Number5673694
Product Code JDI
UDI-Device Identifier07613327012934
UDI-Public07613327012934
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K121308
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2010
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/30/2012
Device Model Number6260-5-332
Device Catalogue Number6260-5-332
Device Lot Number23392101
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/22/2022
Initial Report FDA Received Date09/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/28/2007
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age72 YR
Patient SexFemale
-
-