Catalog Number MX9128
Medical Device Problem Code Obstruction of Flow (2423)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/22/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Date of event: unknown.The date received by manufacturer has been used for this field.There were multiple lot numbers reported to be involved.The information for each lot number is as follows: medical device lot #: 22059113.Medical device expiration date: 11-may-2027, device manufacture date: 06-may-2022.Medical device lot #: 22059057, medical device expiration date: 06-may-2027, device manufacture date: 03-may-2022.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that 1 bd maxguard¿ administration set with needleless y-site each from lots 22059113 and 22059057 were blocked and wouldn't allow fluid past the first hub during use.The following information was provided by the initial reporter: "i am receiving numerous complaints about this iv tubing from end users mainly in our pre-op department.This time it is a different lot #.Staff is reporting that the tubing will not allow fluids to move pass the first hub.No kinks were found.Requiring them to open a new set and exchange out.".
 
Additional Manufacturer Narrative
H6: investigation summary no product or photo was returned by the customer.It was reported that "tubing will not allow fluids to move pass the first hub.No kinks were found".The customer complaint could not be verified due to the product not being returned for failure investigation.Due to no sample being received, an investigation could not be performed and a root cause could not be determined.A device history record review for model mx9128 lot number 22059113 was performed.The search showed that a total of 8,253 units in 1 lot number was built on 12may2022.There were no quality notifications issued for the failure mode reported by the customer during the production build of this set.A device history record review for model mx9128 lot number 22059057 was performed.The search showed that a total of 14,403 units in 1 lot number was built on 11may2022.There were no quality notifications issued for the failure mode reported by the customer during the production build of this set.H3 other text : see h10.
 
Event or Problem Description
It was reported that 1 bd maxguard¿ administration set with needleless y-site each from lots 22059113 and 22059057 were blocked and wouldn't allow fluid past the first hub during use.The following information was provided by the initial reporter: "i am receiving numerous complaints about this iv tubing from end users mainly in our pre-op department.This time it is a different lot #.Staff is reporting that the tubing will not allow fluids to move pass the first hub.No kinks were found.Requiring them to open a new set and exchange out.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD MAXGUARD¿ ADMINISTRATION SET WITH NEEDLELESS Y-SITE
Common Device Name
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15433493
Report Number9616066-2022-01348
Device Sequence Number6071116
Product Code FPA
UDI-Device Identifier10885403237621
UDI-Public10885403237621
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K051499
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue NumberMX9128
Device Lot NumberSEE H10
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/22/2022
Supplement Date Received by Manufacturer10/17/2022
Initial Report FDA Received Date09/15/2022
Supplement Report FDA Received Date10/31/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-