| Model Number |
1011343-40 |
| Medical Device Problem Codes |
Break (1069); Loose or Intermittent Connection (1371)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/19/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned for analysis.The reported loose or intermittent connection was unable to be confirmed due to the condition of the returned device; however there was a noted cracked flush port.The investigation determined the reported/noted difficulties appear to be related to circumstances of the procedure as it is likely that during preparation the syringe was over torqued during connection to the flush port of the device resulting in the noted longitudinal crack running across the entire length of the flush port; thus resulting in the reported loose or intermittent connection.The noted device damages (kinked/separated shaft, wrinkled sheath) likely occurred due to inadvertent mishandling or during packing for return analysis.There is no indication of a product quality issue with respect to design, manufacture or labeling.
|
| |
|
Event or Problem Description
|
|
It was reported that during preparation of a 6-8 x 40mm acculink self-expanding stent system (sess) that saline leaked out of the side of the flushing port where the syringe attaches.The syringe connection was confirmed and no cracks were observed on the sess.The sess was not used in the procedure and was replaced with another acculink sess.There was no patient involvement and no clinically significant delay reported in the procedure.Return device analysis found a break in the shaft (not in two pieces).The account reported they were not aware of the break.No additional information was provided.
|
| |
|
Search Alerts/Recalls
|