• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON & CO. (BROKEN BOW) BD VACUTAINER® TRACE ELEMENT SERUM PLUS BLOOD COLLECTION TUBES; BLOOD SPECIMEN COLLECTION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON & CO. (BROKEN BOW) BD VACUTAINER® TRACE ELEMENT SERUM PLUS BLOOD COLLECTION TUBES; BLOOD SPECIMEN COLLECTION DEVICE Back to Search Results
Catalog Number 368380
Medical Device Problem Code Device Markings/Labelling Problem (2911)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/01/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported when using the bd vacutainer® trace element serum plus blood collection tubes, the device experienced missing label information after being torn.This event occurred once.The following information was provided by the initial reporter.The customer stated: the tube label was found to be torn.
 
Additional Manufacturer Narrative
Investigation summary: bd had not received samples or photos for investigation.Therefore, 100 retention samples from bd inventory were evaluated by visual examination and no issues were observed relating to damaged label as all samples met specifications.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint is unable to be confirmed for the indicated failure mode damaged label.Bd was not able to identify a root cause for the indicated failure mode.Our business team regularly reviews the collected data for identification of emerging trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® TRACE ELEMENT SERUM PLUS BLOOD COLLECTION TUBES
Common Device Name
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BECTON, DICKINSON & CO. (BROKEN BOW)
150 south 1st avenue
broken bow NE 68822
Manufacturer (Section G)
BECTON, DICKINSON & CO. (BROKEN BOW)
150 south 1st avenue
broken bow NE 68822
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15435375
Report Number1917413-2022-00573
Device Sequence Number12161457
Product Code JKA
UDI-Device Identifier00382903683802
UDI-Public00382903683802
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K041071
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date12/31/2022
Device Catalogue Number368380
Device Lot Number1350389
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/01/2022
Initial Report FDA Received Date09/16/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/16/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-