STRYKER NEUROVASCULAR CORK TARGET 360 NANO 1MM X 2CM; DEVICE, NEUROVASCULAR EMBOLIZATION Back to Search Results
Model Number M0035421020
Medical Device Problem Code Premature Separation (4045)
Health Effect - Clinical Code Ruptured Aneurysm (4436)
Date of Event 08/20/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device is not available to the manufacturer.
 
Event or Problem Description
It was reported that during procedure after positioning the subject coil properly into the aneurysm the subject coil didn¿t detach after multiple attempts.When the operator tried to advance the subject coil slightly, the aneurysm ruptured and the subject coil detached in the subarachnoid space.It was also reported that a non-stryker guide catheter had a leak in it that introduced air into the vessel prior to detaching the subject coil.The procedure was stopped right after the aneurysm ruptured and an evd was placed in the patient as a medical intervention.Patient outcome is unknown at this moment.
 
Additional Manufacturer Narrative
Due to the automated mes (manufacturing execution system) system there are controls in the manufacturing process to ensure the product met specifications upon release.The subject device was not available; therefore, a visual inspection as well as a functional evaluation could not be performed.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The reported complaint could not be confirmed, and it could not be definitively determined if the device failed to meet specifications because the product was not returned.While there are a number of potential causes for the reported issue, because review and analysis of available information failed to identify a definitive cause and the device was not returned, an assignable cause of undeterminable will be assigned to reported for 'main coil prematurely detached/separated during use', 'patient aneurysm rupture', and 'main coil failed/unable to detach'.
 
Event or Problem Description
It was reported that during procedure after positioning the subject coil properly into the aneurysm the subject coil didn¿t detach after multiple attempts.When the operator tried to advance the subject coil slightly, the aneurysm ruptured and the subject coil detached in the subarachnoid space.It was also reported that a non-stryker guide catheter had a leak in it that introduced air into the vessel prior to detaching the subject coil.The procedure was stopped right after the aneurysm ruptured and an evd was placed in the patient as a medical intervention.Patient outcome is unknown at this moment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TARGET 360 NANO 1MM X 2CM
Common Device Name
DEVICE, NEUROVASCULAR EMBOLIZATION
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI  NA
Manufacturer (Section G)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI   NA
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key15435759
Report Number3008881809-2022-00453
Device Sequence Number17601280
Product Code HCG
UDI-Device Identifier04546540697967
UDI-Public04546540697967
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K153658
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/06/2024
Device Model NumberM0035421020
Device Catalogue NumberM0035421020
Device Lot Number23276968
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/22/2022
Supplement Date Received by Manufacturer11/03/2022
Initial Report FDA Received Date09/16/2022
Supplement Report FDA Received Date11/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/08/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GUIDE CATHETER (BENCHMARK); INZONE (STRYKER); SL -10 MICROCATHETER (STRYKER)
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient SexUnknown
-
-