HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; VIDEO GASTROSCOPE 2.4C 8.0T 1050L Back to Search Results
Model Number EG-2490K
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/30/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
We checked the returned unit and confirmed that the ccd module fluid damage.Based on the result, we concluded that it was caused due to the fluid damage from the ccd module.In addition, we confirmed that the control body fluid damage, the biopsy inlet t-piece fluid damage, the receptacle fluid damage, the light guide fiber bundle (lcb) fluid damage, the insertion flexible tube (ift) coating damage, the lg prong top broken, the bending rubber perforated, the insertion flexible tube (ift) perforated, the light guide cable buckled, the segment crushed, the lg air/water socket cylinder dirty, the objective lens unit dirty, the lcb distal cover glass dirty, and the lg prong top loose; however, they are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report "hr-rpt-0586(image failure)" and/or the risk analysis results, it was evaluated to submit mdr.
 
Event or Problem Description
The time of event is unknown.There was no report of patient harm.Video image failure(fluid damage).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTAX
Common Device Name
VIDEO GASTROSCOPE 2.4C 8.0T 1050L
Manufacturer (Section D)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA  196-0012
Manufacturer (Section G)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA   196-0012
Manufacturer Contact
william goeller
3 paragon drive
montvale, NJ 07645
MDR Report Key15436041
Report Number9610877-2022-58003
Device Sequence Number14954081
Product Code FDS
UDI-Device Identifier04961333152858
UDI-Public04961333152858
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K131902
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberEG-2490K
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/30/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/30/2022
Initial Report FDA Received Date09/16/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-