• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B. BRAUN MEDICAL INC.; INJECTOR AND SYRINGE, ANGI

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN MEDICAL INC.; INJECTOR AND SYRINGE, ANGI Back to Search Results
Model Number 622510
Medical Device Problem Code Activation Problem (4042)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/02/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).The investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
 
Event or Problem Description
As reported by the user facility: complaint: air was entered when trying to pressurize the balloon.No injury reported.
 
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).One sample was returned for evaluation.The sample was visually and physically evaluated for syringe pressure.There were no cracks visible on the syringe, however, when the pressure test was performed leakage was noticed at the pressure gauge.This could explain why air entered while trying to pressurize the syringe.The reported defect was confirmed.Incidents of this nature are attributed to operator oversight during the assembly of the product.Although our training procedures ensure that all employees are properly trained in their areas of responsibility, an oversight on the part of the operator can attribute to an incident of this nature.Per the sop there was not enough glue dispensed to form a glue "dome" above the rim of the gage port.As a result of this occurrence, a formal awareness training session was conducted with all applicable personnel involved in the assembly and inspection of this product.The purpose of this training was to review the reported incident and to ensure all personnel understand and comply with the established assembly and inspection processes.Review of the discrepancy management system (dsms) database was performed for the reported lot number and no abnormalities or non-conformances were noted during the in process or final product inspection.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NA
Common Device Name
INJECTOR AND SYRINGE, ANGI
Manufacturer (Section D)
B. BRAUN MEDICAL INC.
901 marcon blvd.
allentown PA 18109
Manufacturer Contact
jonathan severino
901 marcon blvd.
allentown, PA 18109
4847197287
MDR Report Key15436846
Report Number2523676-2022-00460
Device Sequence Number12161338
Product Code DXT
UDI-Device Identifier04022495172532
UDI-Public(01)04022495172532
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K902377
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2024
Device Model Number622510
Device Catalogue Number622510
Device Lot Number0061809299
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/06/2022
Supplement Date Received by Manufacturer09/06/2022
Initial Report FDA Received Date09/16/2022
Supplement Report FDA Received Date10/31/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/26/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-