• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC SOFAMOR DANEK USA, INC T2 STRATOSPHERE¿ EXPANDABLE CORPECTOMY SYSTEM; SPINAL VERTEBRAL BODY REPLACEMENT DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC SOFAMOR DANEK USA, INC T2 STRATOSPHERE¿ EXPANDABLE CORPECTOMY SYSTEM; SPINAL VERTEBRAL BODY REPLACEMENT DEVICE Back to Search Results
Model Number 4360258
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Peripheral Nervous Injury (4414)
Date of Event 08/10/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Outcomes attributed to adverse event: other: patient developed a palsy post-operative.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Information was received from healthcare provider (hcp) via a manufacturer representative regarding a patient having anterolateral c orpectomy for burst fracture.Levels implanted: l4 it was reported that that devices broke and were explanted.No fragments were left in the patient. there was no patient symptom reported.Immediately post op, the patient was neurologically intact.However, after undergoing physical therapy, the patient developed some neurological deficit in the left lower extremity.The patient was brought back in for imaging and the instrumentation was intact.It was the surgeons belief that the patient has developed a palsy due to the extended retraction on the psoas muscle.One of the contributing factors in the amount of time the psoas was under retraction was due to the issues with the expandable corpectomy device.These issues increased the amount of time it took to complete the corpectomy by about an hour.All three devices contributed to the delay: the extra time to remove and then attempt to resnap them back on, etc.There were no further complications reported regarding the event.
 
Additional Manufacturer Narrative
H3: product analysis: part # 4360258, lot # ca18j024 visual and optical inspection did not reveal any damage to the end caps.The caps were returned detached from the centerpiece.Functional inspection with a sample centerpiece confirmed the caps were able to be snapped on and removed.The caps function as intended.No fault found.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
T2 STRATOSPHERE¿ EXPANDABLE CORPECTOMY SYSTEM
Common Device Name
SPINAL VERTEBRAL BODY REPLACEMENT DEVICE
Manufacturer (Section D)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
Manufacturer (Section G)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
Manufacturer Contact
glen belmer
1800 pyramid place
memphis, TN 38132
6122713209
MDR Report Key15436851
Report Number1030489-2022-00872
Device Sequence Number14851685
Product Code MQP
UDI-Device Identifier00643169650992
UDI-Public00643169650992
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
K173125
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number4360258
Device Catalogue Number4360258
Device Lot NumberCA18J024
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/24/2022
Supplement Date Received by Manufacturer11/28/2022
Initial Report FDA Received Date09/16/2022
Supplement Report FDA Received Date12/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age72 YR
Patient SexFemale
-
-