• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. BIVONA; TUBE TRACHEOSTOMY AND TUBE CUFF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. BIVONA; TUBE TRACHEOSTOMY AND TUBE CUFF Back to Search Results
Model Number 67PFS50
Medical Device Problem Code Inflation Problem (1310)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the cuffs do not inflate symmetrically.No patient injury was reported.
 
Additional Manufacturer Narrative
A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.Thirteen samples were received for evaluation.Visual and functional testing were performed.Visual inspection found the samples were in used condition.Functional testing found the complaint was not confirmed.No root cause could be determined since the samples successfully passed cuff symmetry test.This issue will be monitored for any increase in occurrence.If the occurrence of this issue increases significantly, additional investigative action will be taken.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms#: 617147.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIVONA
Common Device Name
TUBE TRACHEOSTOMY AND TUBE CUFF
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
5700 west 23rd ave
gary IN 46406
Manufacturer (Section G)
SMITHS MEDICAL ASD, INC.
5700 west 23rd ave
gary IN 46406
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15436913
Report Number3012307300-2022-19396
Device Sequence Number5673267
Product Code JOH
UDI-Device Identifier10351688518644
UDI-Public10351688518644
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K923878
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date09/19/2025
Device Model Number67PFS50
Device Catalogue Number67PFS50
Device Lot Number4053704
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/03/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/10/2022
Initial Report FDA Received Date09/16/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-