Model Number 67PFSS50
Medical Device Problem Code Inflation Problem (1310)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/09/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Samples received: one (1) samples was received in used conditions without its original packaging and without its certificate of decontamination.Visual inspection: visual inspection was performed at 12? under normal lighting to received unit, in order to detect any damage on the cuff, airline or pilot balloon.No damaged could be detected.Functional test: sample was filled with air to see if there was any functional problem.When inflating the sample with air, it was seen that cuff inflated uniformly.The balloon was manipulated and the whole cuff inflated.After the whole cuff inflated, it was deflated and inflated 4 times, all the times the cuff inflated completely.A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
 
Event or Problem Description
It was reported a new trach was put on the patient when the parent noticed that the patient started to desat.The patient then had to be bagged with a resus bag.They removed the trach and saw that it was not fully inflated.The patient's mother then tested the backup trach they had and found it would not fully inflate either so mom put the old trach back in the patient.The mother did not want to do this as the old trach has mold in it.No patient injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIVONA
Common Device Name
TUBE TRACHEOSTOMY AND TUBE CUFF
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
5700 west 23rd ave
gary IN 46406
Manufacturer (Section G)
SMITHS MEDICAL ASD, INC.
5700 west 23rd ave
gary IN 46406
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15437013
Report Number3012307300-2022-19405
Device Sequence Number5836303
Product Code JOH
UDI-Device Identifier10351688518712
UDI-Public10351688518712
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K923878
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number67PFSS50
Device Catalogue Number67PFSS50
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/25/2020
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/10/2022
Initial Report FDA Received Date09/16/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-