Samples received: one (1) samples was received in used conditions without its original packaging and without its certificate of decontamination.Visual inspection: visual inspection was performed at 12? under normal lighting to received unit, in order to detect any damage on the cuff, airline or pilot balloon.No damaged could be detected.Functional test: sample was filled with air to see if there was any functional problem.When inflating the sample with air, it was seen that cuff inflated uniformly.The balloon was manipulated and the whole cuff inflated.After the whole cuff inflated, it was deflated and inflated 4 times, all the times the cuff inflated completely.A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
|