SYNTHES GMBH BURR ATTACHMENT-MED-ANGLED F/PEN DRIVE; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
|
Back to Search Results |
|
| Catalog Number |
05.001.049 |
| Medical Device Problem Codes |
Insufficient Heating (1287); Complete Loss of Power (4015)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported by switzerland that during service and evaluation, it was determined that the burr attachment device generated heat.It was further observed that the bearing was worn.It was further determined that the device failed pretest for check tool coupling.It was noted in the service order that during pre-surgery, it was discovered that the device did not work anymore.It was not reported if there were any delays in the surgical procedure or if a spare device was available for use.There was no patient involvement reported.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.Device evaluation: the actual device was returned for evaluation.During repair, an evaluation was performed, and it was determined that the reported condition of the device not working was not confirmed.Therefore, an assignable root cause was not determined.However, the device generating heat identified during service and evaluation was confirmed.The assignable root cause was determined to be due to component failure from normal wear.Udi #: (b)(4).
|
| |
|
Search Alerts/Recalls
|
|
|