| Model Number |
LDA220/65 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Death
|
|
Event or Problem Description
|
|
Related manufacturer report number: 2017865-2022-38361 and 2017865-2022-38363.It was reported that the patient had deceased.The cause of death was cardiac arrhythmia, dilated cardiomyopathy and cardiomegaly.There is no known allegation from a health care professional that suggests that the death was device related.No additional information was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The lead was returned due to patient deceased.As received, a complete lead was returned in one piece for analysis.Visual inspection of the lead found no anomalies with the exception of damages consistent with procedural damage.X-ray examination found the broken and separated svc cables at the connector region consistent with procedural damage.Electrical testing found the open continuity svc cables.
|
| |
|