• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MPRI ATTAIN ABILITY STRAIGHT MRI SURESCAN; DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) P

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MPRI ATTAIN ABILITY STRAIGHT MRI SURESCAN; DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) P Back to Search Results
Model Number 439688
Medical Device Problem Codes Failure to Capture (1081); High Capture Threshold (3266)
Health Effect - Clinical Codes Presyncope (4410); Syncope/Fainting (4411)
Date of Event 07/06/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Concomitant medical products: dtpb2d4 crt-d implanted: (b)(6) 2022.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the patient experience syncope and pre-syncope conditions.It was noted that the left ventricular (lv) lead exhibited high thresholds with suspected capture issue noted.The lv lead remains in use.No further patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTAIN ABILITY STRAIGHT MRI SURESCAN
Common Device Name
DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) P
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key15438236
Report Number2649622-2022-18540
Device Sequence Number12180816
Product Code OJX
UDI-Device Identifier00643169860230
UDI-Public00643169860230
Combination Product (Y/N)Y
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
P080006
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date09/10/2020
Device Model Number439688
Device Catalogue Number439688
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Fifteen-Day
Initial Date Received by Manufacturer 09/10/2022
Supplement Date Received by Manufacturer10/27/2022
Initial Report FDA Received Date09/16/2022
Supplement Report FDA Received Date10/27/2022
Date Device Manufactured10/10/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
6935M55 LEAD, 5076-45 LEAD
Outcome Attributed to Adverse Event Other; Life Threatening;
Patient Age69 YR
Patient SexUnknown
-
-