• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 DECATHLON CATHETER; DIALYSIS CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 DECATHLON CATHETER; DIALYSIS CATHETER Back to Search Results
Model Number DF19SH24K
Medical Device Problem Codes Fracture (1260); Failure to Advance (2524)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/24/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.However, photos were provided for review.The investigation of the reported event is currently underway.Expiry date: 07/2023.The information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.Device pending return.
 
Event or Problem Description
It was reported that during a dialysis catheter placement procedure, the device allegedly failed to advance.It was further reported that the portion of the device was broken and pinching on the chest of the patient.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: one tunneler with a loaded protective sheath were returned for evaluation.Gross visual and microscopic visual evaluations were performed.Although the sample was returned for evaluation, four electronic photos were provided for review.The investigation is confirmed for the reported fracture issue, as the protective sheath loaded on the tunneler was noted to have a partial circumferential break.However, the investigation is inconclusive for the reported failure to advance issue, as the exact circumstances at the time of the reported event cannot be verified, and the reported event could not be reproduced in the lab.A definitive root cause could not be determined based upon the available information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 07/2023).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that during a dialysis catheter placement procedure, the device allegedly failed to advance.It was further reported that the portion of the device was broken and pinching on the chest of the patient.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DECATHLON CATHETER
Common Device Name
DIALYSIS CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key15439635
Report Number3006260740-2022-03636
Device Sequence Number12255575
Product Code MSD
UDI-Device Identifier00801741013942
UDI-Public(01)00801741013942
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K033294
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberDF19SH24K
Device Catalogue NumberDF19SH24K
Device Lot NumberREGS2888
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/26/2022
Supplement Date Received by Manufacturer11/01/2022
Initial Report FDA Received Date09/16/2022
Supplement Report FDA Received Date11/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age60 YR
Patient SexFemale
-
-