MAUDE Adverse Event Report: COOK INC FLEXOR RAABE GUIDING SHEATH; DYB INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC FLEXOR RAABE GUIDING SHEATH; DYB INTRODUCER, CATHETER Back to Search Results
Model Number G12266
Medical Device Problem Codes Break (1069); Unraveled Material (1664)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 08/31/2022
Type of Reportable Event Serious Injury
Event or Problem Description
As reported, during a procedure within the distal superficial femoral artery, a flexor raabe guiding sheath separated in half upon removal from the patient.The patient was taken to surgery and the separated portion of the device was removed.The patient reportedly "did fine".Photos of the returned complaint device show unraveling and separation of the sheath.Additional information has been requested.
 
Additional Manufacturer Narrative
(b)(4).This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
Additional information was received (b)(6) 2022.Access was obtained in the right common femoral artery for a contralateral approach, up and over the aortic bifurcation, to target the left superficial femoral artery.Another manufacturer's balloon was used through the sheath during the procedure.The dilator was not in place when the sheath separated mid-shaft.A vascular surgeon removed the distal half of the sheath from the patient during a cut-down of the contralateral common femoral artery.The patient was hospitalized overnight and was discharged the following day.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Additional information: b2, b5, h6 (annex f), d10 this report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unchanged, or unavailable.Additional/corrected information: b5, e4 summary of event: as reported, during a procedure within the distal superficial femoral artery, a flexor raabe guiding sheath separated in half upon removal from the patient.The patient was taken to surgery and the separated portion of the device was removed.The patient reportedly "did fine".Photos of the returned complaint device show unraveling and separation of the sheath.Additional information was received (b)(6) 2022.Access was obtained in the right common femoral artery for a contralateral approach, up and over the aortic bifurcation, to target the left superficial femoral artery.Another manufacturer's balloon was used through the sheath during the procedure.The dilator was not in place when the sheath separated mid-shaft.A vascular surgeon removed the distal half of the sheath from the patient during a cut-down of the contralateral common femoral artery.The patient was hospitalized overnight and was discharged the following day.Additional information was received (b)(6) 2022.The distal half of the sheath was removed with a snare, via access in the contralateral groin, by a vascular surgeon during a separate procedure.There were no adverse effects to the patient.Investigation evaluation: reviews of the complaint history, device history record, instructions for use (ifu), and quality control procedures were conducted during the investigation.A visual inspection of the complaint device was also conducted.The complaint device was returned to cook for investigation.The sheath was separated at 42-centimeters from the hub.A document-based investigation evaluation was performed.No related non-conformances were found, and there have been no other reported complaints for this lot number.The product ifu warns that reinsertion of the dilator prior to removal of the sheath ¿increases the strength of the sheath and lessens the risk of device separation¿ and suggests insertion of the dilator prior to removal if resistance is encountered or anticipated during withdrawal of the device.The ifu also cautions ¿in order to ensure device compatibility, choose a sheath size large enough to accommodate the maximum outer diameter of any devices that will be placed through the sheath.All interventional or diagnostic instruments used with this product should move freely through the valve and sheath to avoid damage.¿ the information provided upon review of the dmr, dhr, ifu, and investigation of the returned device suggests that there is evidence the device was manufactured to specification.There is no evidence of non-conforming devices in-house or in the field.Cook has concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.Based on the information provided and the results of the investigation, cook was unable to determine a definitive cause for this event.The risk analysis for this failure mode was reviewed and no additional escalation was required.There is currently a capa investigation open to investigate this product failure.The appropriate personnel have been notified and cook will continue to monitor for similar events.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
Additional information was received (b)(6) 2022.The distal half of the sheath was removed with a snare, via access in the contralateral groin, by a vascular surgeon during a separate procedure.There were no adverse effects to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXOR RAABE GUIDING SHEATH
Common Device Name
DYB INTRODUCER, CATHETER
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC.
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key15439745
Report Number1820334-2022-01492
Device Sequence Number12158166
Product Code DYB
UDI-Device Identifier00827002122666
UDI-Public(01)00827002122666(17)250208(10)14521128
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K142829
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/08/2025
Device Model NumberG12266
Device Catalogue NumberKCFW-6.0-38-90-RB-RAABE
Device Lot Number14521128
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/13/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 08/31/2022
Supplement Date Received by Manufacturer10/06/2022
11/16/2022
Initial Report FDA Received Date09/16/2022
Supplement Report FDA Received Date11/01/2022
12/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/08/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SHOCKWAVE BALLOON
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexUnknown
-
-