| Model Number |
CLV-S200-IR |
| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has not been received to date.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time, however, if additional information becomes available, this report will be supplemented accordingly.
|
| |
|
Event or Problem Description
|
|
The user facility returned an asset visera elite ii xenon light source to olympus to the service center.The user facility did not report a malfunction with the device.Upon inspection and testing of the returned unit, the device does not power on.There was no patient involvement associated with this event.This medical device report (mdr) is being submitted to capture the reportable malfunction found during evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and device evaluation.Please see updates to d8, d9, h4, h6 and h10.The device was returned to olympus for inspection, and the reportable malfunction was confirmed.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 5 years since the subject device was manufactured.Based on the results of the investigation, it¿s likely the device did not power on due to a damaged power u harness switch.The root cause of this event was unable to be identified.Olympus will continue to monitor field performance for this device.
|
| |
|