| Model Number |
SRL-510 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
07/27/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side rupture.The device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: rupture: observed one opening in device on the anterior side assessed as fold flaw opening.Lump/nodule-ndr: not applicable as the event is not related to the device.Additional observations.Crease fold observed on the device.Non penetrating nick (stress marks).
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side rupture.Patient's spouse reported "a lump on the right side that was tested and benign"- not device related.Device has been explanted and replaced.
|
| |
|
Event or Problem Description
|
|
Healthcare profesional reported right side rupture.The device remains implanted.
|
| |
|