COOK INC FLEXOR ANSEL GUIDING SHEATH; DYB INTRODUCER, CATHETER Back to Search Results
Model Number G44155
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/30/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown or unavailable.Name and address- phone: (b)(6).This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
As reported, during preparation for a left leg angioplasty, the tip of a flexor ansel guiding sheath split.The dilator was inserted into the sheath, and both the sheath and dilator were attempted to be advanced through another manufacturer's 7fr sheath.The tip of the sheath was then noted to be split, and a new sheath was opened to continue the procedure.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Additional Manufacturer Narrative
B1, h1: upon return and evaluation of the device on 04oct2022, the sheath tip was folded inward, but was not split as originally reported by the customer.Per a search of risk documentation and previous complaint data, there is no evidence to suggest that this device failure would be likely to cause or contribute to a death or a serious injury if the failure were to recur.Furthermore, there is no evidence that the device caused or contributed to a serious injury, as no life-threatening or permanently impairing injury was alleged, nor was intervention taken as a result of the device failure which would be required to prevent permanent impairment or damage to the patient.As such, the event is not reportable under fda 21 cfr part 803 as it does not meet the criteria for a death, serious injury, or reportable product malfunction.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
Upon return and evaluation of the device, the tip of the sheath was not split as reported by the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXOR ANSEL GUIDING SHEATH
Common Device Name
DYB INTRODUCER, CATHETER
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key15440082
Report Number1820334-2022-01494
Device Sequence Number12161228
Product Code DYB
UDI-Device Identifier00827002441552
UDI-Public(01)00827002441552(17)250415(10)14669643
Combination Product (Y/N)N
PMA/510(K) Number
K142829
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/15/2025
Device Model NumberG44155
Device Catalogue NumberKCFW-7.0-18/38-45-RB-ANL0-HC
Device Lot Number14669643
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 08/31/2022
Supplement Date Received by Manufacturer10/04/2022
Initial Report FDA Received Date09/16/2022
Supplement Report FDA Received Date10/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/15/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CORDIS 7FR BRIGHT TIP SHEATH.; TERUMO 0.18 WIRE.
Patient SexUnknown
-
-