Model Number 0684-00-0479-02
Medical Device Problem Code Difficult to Insert (1316)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/26/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has not been returned to the manufacturer so we are unable to complete an evaluation.If provided we will send a supplemental report with our additional findings.Complaint record id # (b)(4).
 
Event or Problem Description
It was reported that the intra-aortic balloon (iab) could not be inserted through the sheath.There was no patient harm or adverse event reported.
 
Additional Manufacturer Narrative
Complaint record id # (b)(4).
 
Event or Problem Description
It was reported that the intra-aortic balloon (iab) could not be inserted through the sheath.A new iab was successfully inserted to continue therapy.There was no patient harm or adverse event reported.
 
Additional Manufacturer Narrative
The device has not been returned to the manufacturer so we are unable to complete an evaluation.Reference complaint (b)(4).
 
Event or Problem Description
N/a.
 
Additional Manufacturer Narrative
The product was returned with the membrane completely unfolded and traces of blood on the exterior.The sheath, dilator, step dilator and pressure tubing were also returned.The returned sheath was measured and found to be dimensionally within specification.A laboratory insertion test was unable to be performed due to the membrane being unfurled.An underwater leak test of the balloon, catheter, y-fitting and extracorporeal tubing was performed and no leaks were detected.A leak may impact the ability to maintain vacuum.We are unable to confirm the reported difficulty during insertion because of the returned condition of the catheter and we are unable to mimic the clinical setting.A lot history record review was completed for the reported product.No non-conformances were found that are considered to be related to the event.Reference complaint #(b)(4).
 
Event or Problem Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LINEAR 7.5 FR. 34CC IAB
Common Device Name
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - FAIRFIELD
15 law drive
fairfield NJ
Manufacturer (Section G)
DATASCOPE CORP. - FAIRFIELD
15 law drive
fairfield NJ
Manufacturer Contact
brian schaeffer
15 law drive
fairfield, NJ 
MDR Report Key15440892
Report Number2248146-2022-00725
Device Sequence Number14848246
Product Code DSP
UDI-Device Identifier10607567106540
UDI-Public10607567106540
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K041281
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 03/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/12/2024
Device Model Number0684-00-0479-02
Device Catalogue Number0684-00-0474
Device Lot Number3000198520
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/27/2022
Supplement Date Received by Manufacturer09/21/2022
01/20/2023
03/13/2023
Initial Report FDA Received Date09/16/2022
Supplement Report FDA Received Date10/07/2022
01/20/2023
03/13/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/13/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
-
-