| Model Number |
0684-00-0479-02 |
| Medical Device Problem Code |
Difficult to Insert (1316)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/26/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has not been returned to the manufacturer so we are unable to complete an evaluation.If provided we will send a supplemental report with our additional findings.Complaint record id # (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the intra-aortic balloon (iab) could not be inserted through the sheath.There was no patient harm or adverse event reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint record id # (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the intra-aortic balloon (iab) could not be inserted through the sheath.A new iab was successfully inserted to continue therapy.There was no patient harm or adverse event reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has not been returned to the manufacturer so we are unable to complete an evaluation.Reference complaint (b)(4).
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|
Additional Manufacturer Narrative
|
|
The product was returned with the membrane completely unfolded and traces of blood on the exterior.The sheath, dilator, step dilator and pressure tubing were also returned.The returned sheath was measured and found to be dimensionally within specification.A laboratory insertion test was unable to be performed due to the membrane being unfurled.An underwater leak test of the balloon, catheter, y-fitting and extracorporeal tubing was performed and no leaks were detected.A leak may impact the ability to maintain vacuum.We are unable to confirm the reported difficulty during insertion because of the returned condition of the catheter and we are unable to mimic the clinical setting.A lot history record review was completed for the reported product.No non-conformances were found that are considered to be related to the event.Reference complaint #(b)(4).
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|