| Model Number |
IPN037417 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/13/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Qn#(b)(4).
|
| |
|
Event or Problem Description
|
|
The complaint is reported as "the female luer lock hub on the end of the lumen came off.The lumen had to be clamped and the catheter was replaced.".No patient harm was reported.The patient's condition is reported as fine.
|
| |
|
Additional Manufacturer Narrative
|
|
Qn# (b)(4).The actual device was not returned; however, the customer provided one photo for analysis.The complaint of extension line and luer hub separation was able to be confirmed by the photo.A complete visual inspection could not be performed as no sample was returned for analysis.A device history record review was performed, and no relevant findings were identified.The instructions for use (ifu) provided with this kit warns the user, "warning: open slide clamp prior to infusion through lumen to reduce risk of damage to extension line from excessive pressure." the customer report of extension line and luer hub separation was confirmed by visual inspection of the customer supplied photo.However, full complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed, and no relevant findings were identified.Without the device to evaluate, the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
|
| |
|
Event or Problem Description
|
|
The complaint is reported as "the female luer lock hub on the end of the lumen came off.The lumen had to be clamped and the catheter was replaced.".No patient harm was reported.The patient's condition is reported as fine.
|
| |
|
Search Alerts/Recalls
|