MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II; PUMP, INFUSION, IMPLANTED, PROGRAMMABLE
|
Back to Search Results |
|
Model Number 8637-20 |
Device Problem
Connection Problem (2900)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/05/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Other relevant device(s) are: product id: 8781, serial/lot #: (b)(4), ubd: 27-oct-2023, udi#: (b)(4).Product type catheter medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Information was received from a foreign healthcare provider (hcp) via a distributor regarding a patient who was receiving an unknown drug via an implantable pump.On (b)(6) 2022, it was reported that, while attempting to connect the pump catheter to the spinal catheter, the collet had already been engaged and the catheter could not be inserted into the collet connector.The connector was considered defective as the surgery could not be discontinued, a different accessory kit was used.No patient symptoms were reported.
|
|
Event Description
|
Additional information was received from a foreign healthcare provider via a distributor on 2022-(b)(6).Surgery at the time of the occurrence was regarding implantation of a new pump on 2022-(b)(6); there was no catheter replacement, etc.The drug administered via the pump was gabalon intrathecal.The male patient¿s age was 57 years old at the time of the event.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
H3: the returned pin connector was subjected to a series of standard tests that include but is not limited to visual inspection, patency testing, and pressure testing.Analysis identified that the catheter was not fully seated onto the connector and the collet was locked in place medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|