MAUDE Adverse Event Report: ARROW INTERNATIONAL LLC ULTRAFLEX IAB: 7.5FR 40CC; SYSTEM, BALLOON, INTRA-AORTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL LLC ULTRAFLEX IAB: 7.5FR 40CC; SYSTEM, BALLOON, INTRA-AORTIC Back to Search Results
Model Number IPN039788
Medical Device Problem Code Gas/Air Leak (2946)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/17/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "after the catheter was implanted during the operation, the machine alarmed (gas leak alarm).After the catheter was taken out, the catheter was found leakage".Additional information stated that the catheter was replaced with a new iab.The new iab was inserted at the same insertion site.Patient condition reported as "fine".
 
Additional Manufacturer Narrative
Qn# (b)(4).
 
Event or Problem Description
It was reported that "after the catheter was implanted during the operation, the machine alarmed (gas leak alarm).After the catheter was taken out, the catheter was found leakage".Additional information stated that the catheter was replaced with a new iab.The new iab was inserted at the same insertion site.Patient condition reported as "fine".
 
Additional Manufacturer Narrative
(b)(4).Returned for investigation was a 40cc 7.5fr ultra flex intra-aortic balloon catheter (iabc) with the original packaging box that does not match the serial number on the returned sample.Upon return, the one-way valve was tethered to the short driveline tubing.The bladder was fully unwrapped.Dried blood was noted on the exterior surfaces of the returned iabc; no obvious blood was noted within the helium pathway.No visual damage or abnormalities were noted to the returned sample.The one-way valve was tested and passed.A vacuum was pulled on the one-way valve, and it held for at least 1 minute and then 30 seconds five separate times in accordance with the quality system document.The iabc central lumen was aspirated and flushed.No abnormalities or debris were noted.The iabc was leak tested in accordance with testing methods from the manufacturing procedure.A leak was immediately noted and located around the bifurcate bushing bond.Upon microscopic inspection, the leak occurred from the adhesive bond at the bifurcate bushing.No other leaks were detected.A guidewire was back loaded through the iabc distal tip.No resistance was noted; the guidewire was able to advance through the central lumen.No blood or debris was noted.The guidewire was front loaded through the iabc luer.No resistance was noted; the guidewire was able to advance through the central lumen.No blood or debris were noted.A device history record (dhr) review was conducted for the lot number with no relevant findings.The device passed all manufacturing specifications prior to release.The reported complaint for iab helium loss alarm is confirmed.During the complaint investigation, an external leak is confirmed from the intra-aortic balloon catheter (iabc) bifurcate bushing adhesive.Based on a review of the device history record (dhr), the product met specification upon release; however, the specifications were not met during the complaint investigation due to the external leak from the iabc bifurcate bushing adhesive.The root cause is manufacturing related.A non-conformance has been initiated to further investigate the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRAFLEX IAB: 7.5FR 40CC
Common Device Name
SYSTEM, BALLOON, INTRA-AORTIC
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL INC.
16 elizabeth drive
chelmsford MA 01824
Manufacturer Contact
bryanna connelly
3015 carrington mill blvd
morrisville 27560
MDR Report Key15442705
Report Number3010532612-2022-00355
Device Sequence Number5672919
Product Code DSP
UDI-Device Identifier10801902145731
UDI-Public10801902145731
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K000729
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date03/31/2024
Device Model NumberIPN039788
Device Catalogue NumberIAB-06840-U
Device Lot Number18F22D0002
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/29/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/19/2022
Supplement Date Received by Manufacturer10/07/2022
Initial Report FDA Received Date09/16/2022
Supplement Report FDA Received Date10/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/04/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NONE REPORTED; NONE REPORTED
Patient SexUnknown
-
-