• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIFESCAN EUROPE GMBH OT VERIO 2 METER; GLUCOSE MONITORING SYS/KIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIFESCAN EUROPE GMBH OT VERIO 2 METER; GLUCOSE MONITORING SYS/KIT Back to Search Results
Model Number 022-659
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Codes Hypoglycemia (1912); Diaphoresis (2452); Shaking/Tremors (2515)
Date of Event 09/08/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Similar complaints for this issue were trended including the reported meter.It was concluded that the number of complaints for the meter did not breach thresholds indicative of a systemic issue.In addition, similar complaints for this issue were trended for the test strip lot.It was concluded that the number of complaints for the lot did not breach thresholds indicative of a systemic issue.
 
Event or Problem Description
On (b)(6) 2022, the lay-user/patient contacted lifescan (lfs) united states, alleging that their onetouch verio2 meter read inaccurately high compared to their feelings and/or normal readings.The complaint was classified based on the customer care agent (cca) documentation and on further information obtained by the medical surveillance specialist after reviewing the call recording.The patient reported that the alleged meter inaccuracy began 3 weeks prior to contacting lfs.The patient reported obtaining what they felt were inaccurate high blood glucose readings of ¿145, 120, 80, 109, 108, 110, 134 and 147 mg/dl¿ with the subject meter on unspecified dates/times.The patient manages their diabetes with oral medication (metformin) and they continued to follow their usual diabetes management regimen in response to the alleged issue.The patient reported that on (b)(6) 2022, at 5:30 pm, they experienced a ¿low sugar¿ with symptoms of ¿sweating and shaking¿.The patient claimed they self-treated with orange juice.No other device was used for testing.At the time of troubleshooting, the cca confirmed the unit of measure was set correctly on the subject meter.The cca confirmed the test strip vial was intact and the test strips had been stored correct and were not expired or opened past their discard date.The cca noted the patient did not have control solution available to perform a quality control test.Replacement products were sent to the patient.This complaint is being reported because the patient reportedly developed symptoms suggestive of a serious injury adverse event after the alleged inaccuracy issue began.The subject meter could not be ruled out as a cause or contributor to the event.
 
Additional Manufacturer Narrative
The lay user/patient¿s meter has been returned and evaluated by lifescan product analysis with the following findings: the meter has passed performance testing with no faults found.The reported issue could not be confirmed.In addition, a device history record review was performed on the subject meter lot.The review did not identify anything that could adversely impact product performance or function.If lifescan obtains additional information regarding this complaint, a follow-up report will be submitted.At this time, lifescan considers this matter closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OT VERIO 2 METER
Common Device Name
GLUCOSE MONITORING SYS/KIT
Manufacturer (Section D)
LIFESCAN EUROPE GMBH
gubelstrasse 34
zug 6300
SZ  6300
Manufacturer (Section G)
LIFESCAN SCOTLAND
beechwood park north
inverness IV2 3 ED
UK   IV2 3ED
Manufacturer Contact
simon palmer
beechwood park north
inverness IV2 3-ED
UK   IV2 3ED
1463383679
MDR Report Key15442745
Report Number3008382007-2022-04343
Device Sequence Number5582427
Product Code NBW
UDI-Device Identifier00353885008730
UDI-Public00353885008730
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K131363
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 03/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number022-659
Device Lot Number4857083
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/09/2022
Supplement Date Received by Manufacturer01/25/2023
Initial Report FDA Received Date09/16/2022
Supplement Report FDA Received Date03/23/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening;
Patient SexUnknown
-
-